medical Class II Updated 2020-08-12

Siemens Healthcare Diagnostics, Inc. recalls DCA Vantage Handheld Barcode Scanner - Zebra Model - Model #

Recalled Product

DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US


Hazard / Issue

If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: DCA Vantage Handheld Barcode Scanner #11306432 Model: DS4308 - HC0015BZZWW
View official government recall

Looking for safe alternatives?