Terumo Cardiovascular Systems Corporation recalls CDI Blood parameter monitoring system 500 H/SAT, Part No. 14
Recalled Product
CDI Blood parameter monitoring system 500 H/SAT, Part No. 145883 Product Usage: The CDI System 500 provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. In addition, calculated values of base excess, bicarbonate, oxygen saturation, and oxygen consumption may also be provided. These parameters are displayed at either actual temperature or adjusted to 37¿C. For documentation purposes, the system 500s integral printer provides a hard copy of displayed parameters.
Hazard / Issue
H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.