Boston Scientific Corporation recalls FARADRIVE Steerable Sheath Clear IDE, used to facilitate acc
See all recalls from Boston Scientific Corporation →Recalled Product
FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and into the chambers of the heart, Material Number 21C602, IFU Reference Number 51636751-01
Hazard / Issue
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Issued by
FDA
✅ What Should You Do?
- 1Contact your healthcare provider immediately to discuss this recall.
- 2Stop using the device if it is safe to do so, and contact your provider.
- 3Follow the manufacturer's remedy instructions provided in the recall notice.
- 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
Looking for a safe alternative?
Search Amazon for FARADRIVE Steerable Sheath Clear →If you were harmed by this recalled product, Top Class Actions lists active class action lawsuits and settlements that may apply to you.
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.