medical Class II Updated 2018-09-26

Degania Silicone, Ltd. recalls Temperature Sensor Catheter 8FR, Catalogue Number 1022011008

Recalled Product

Temperature Sensor Catheter 8FR, Catalogue Number 102201100863BI Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.


Hazard / Issue

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: V17002337, V17003078, V17016095, V17020364
View official government recall