medical Class II Updated 2019-10-09

Beckman Coulter Inc. recalls The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot

Recalled Product

The Access Cortisol Reagent. REF (Catalog Number) 33600, Lot 831755. A paramagnetic particle, chemiluminscent immunoassay. Product Usage: The assay is a paramagnetic particle, chemiluminescent immunoassay for the quanitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.


Hazard / Issue

Cross contamination may have occurred between wells for the reagent pack. As a result of this defect, patient results may be negatively impacted (falsely decreased) by up to 16%).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number 831755.