medical Class II Updated 2018-09-26

Degania Silicone, Ltd. recalls Foley Catheter with temperature sensor 400TM 18FR, Catalogue

Recalled Product

Foley Catheter with temperature sensor 400TM 18FR, Catalogue Number 102201101880TY Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.


Hazard / Issue

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: S17002333, S17002452, S17002892, S17003578, S17005045, S17003306, S17005046, S17007065, S17005298, S17000279, S17000280, S17000837, S17007171, S18001346, S17000333
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →