medical Class II Updated 2020-08-12

Covidien Llc recalls Beacon EUS Access System Preloaded Access 135degree Needle

Recalled Product

Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.


Hazard / Issue

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Issued by

FDA

Distribution: world wide distribution
Lot/Code Info: Model # DSA-135-01 UDI: 10884521663527 Lot #: F2510371X, F2510861X, F2512341X, F2512383X, F2512493X, F2513453X
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →