medical Class II Updated 2020-08-19

Ion Beam Applications S.A. recalls Proteus 235 The Proton Therapy System - Proteus 235 (brand

Recalled Product

Proteus 235 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.


Hazard / Issue

Signature from the user is necessary to proceed with specific actions in the Proton Therapy System (PTS). IBA became aware that the PTS does not accept user names with more than ten characters. It is an issue when the user has no other choice than resuming an aborted treatment field based on the overall delivered dose displayed on the Dose Counter Electronic Unit (DCEU). User s signature is required to perform this action. If the signature contains more than ten characters, the user will not be able to complete the aborted treatment field. An additional issue applies to Electronic Medical Record (EMR) centric sites and may increase the probability of not being able to complete an aborted treatment field. It is not possible, in a new session, to resume from a local partial archive if the Patient Positioning System (PPS) position has changed. If the user captures the PPS position at every session in the Oncology Information System (OIS) just after the setup process, the prescribed PPS position is changed for the next session in the OIS. This includes the partial continuation session. Therefore, when comparing the prescribed PPS position between OIS and local database, the PTS sees a difference and rejects the local partial archive. This problem forces the user to resume the interrupted irradiation based on the overall delivered dose displayed on the DCEU instead of resuming from the full details of the interrupted beam.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Number: Manufacture date SAT.140 (US): manufactured May 2019 SBF.112 (EU): manufactured October 2019 SBF.113 (US): manufactured June 2019 SBF.117 (EU): manufactured November 2019
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →