medical Class II Updated 2018-09-26

Degania Silicone, Ltd. recalls Thermistor Foley catheter, nonsterile, Catalogue Number 1022

Recalled Product

Thermistor Foley catheter, nonsterile, Catalogue Number 102201101863MD Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.


Hazard / Issue

The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: V17011921, V17026466, V17026467, V17026862
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →