medical Class II Updated 2020-08-19

Shanghai United Imaging Healthcare Co., Ltd. recalls uMI 550 PET/CT Diagnostic Imaging System - Product Usage: Th

Recalled Product

uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.


Hazard / Issue

1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial No.s 200017, 200023, 200024 and 200036.
View official government recall