Philips North America, LLC recalls Phillips Charging Station, Product #: 989803191021 - Product
Recalled Product
Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR400 MRI Patient Monitoring System (Model MR400) is intended to monitor vital signs for patients undergoing MRI procedures and to provide signals for synchronization for the MRI scanner.
Hazard / Issue
A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.