medical Class II Updated 2018-08-22

ICU Medical de Mexico, S.A. de C.V. recalls 10" Smallbore Ext Set w/6-Port NanoClave Manifold (Orange, R

Recalled Product

10" Smallbore Ext Set w/6-Port NanoClave Manifold (Orange, Red, Blue, Purple, Yellow, Green Rings), NanoClave (Light Blue Ring), Check Valve, Clamp, Rotating Luer.


Hazard / Issue

ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.

Issued by

FDA

Affected States: AZ, MA, NC, OH, TX, VA, WA
Lot/Code Info: Item Number AM6128. a. Lot Number 3423330, UDI Number (01) 1 0887709 08696 5 (17) 220401 (30) 50 (10) 3423330, Expiration Date 04/01/2022; b. Lot Number 3476737, UDI Number (01) 1 0887709 08696 5 (17) 220601 (30) 50 (10) 3476737, Expiration Date 06/01/2022
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →