medical Class II Updated 2018-08-29

Siemens Medical Solutions USA, Inc recalls 3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3

Recalled Product

3D TOP Ceiling Stand, model no. 3070039 3074486 3070013 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.


Hazard / Issue

In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 3070039: S/N 2271 to 2793 3074486: S/N 1812 to 3036 3070013: S/N 3967 to 5042 3070021: S/N 3852 to 4719