medical Class II Updated 2018-08-29

Philips Electronics North America Corporation recalls Philips SureSigns VSV (Vital Signs Viewer), 863067 Indica

Recalled Product

Philips SureSigns VSV (Vital Signs Viewer), 863067 Indicated for central monitoring of multiple adult, pediatric and neonatal patients, and where the clinician decides to monitor cardiac arrhythmia of adult, pediatric and neonatal patients to gain information for the treatment, to monitor adequacy of treatment, or to exclude causes of symptoms.


Hazard / Issue

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited. These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles. Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery. An overheated battery may in turn cause the device case to overheat and possibly melt or cause the device to ignite, which can cause injury to a patient, nearby users, or cause damage to property.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI# VSV (863067): 00884838009578
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →