medical Class II Updated 2020-09-02

Neurovision Medical Products Inc recalls Cobra x5 2-Ch EMG ET Tube, 6.0mm, REF: LTE700DCS-5, Rx Only,

Recalled Product

Cobra x5 2-Ch EMG ET Tube, 6.0mm, REF: LTE700DCS-5, Rx Only, Sterile EO UDI: B006LTE700DCS52/$$7062618C - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.


Hazard / Issue

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Kit Lots: 022119A 050219A 051419A 060519A 062119B 081919B 082019C 082619B 082719C 061720B 061720D 120919D