medical MODERATE Updated 2026-08-12

Contego Medical, Inc recalls Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloo

See all recalls from Contego Medical, Inc →

Recalled Product

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI-DI/Lot: NG-0630-140-2/00851616007583/Z2460578D,Z2460740B, Z2460872C, Z2460909F, Z2460948D, Z2460957Q; NG-0640-140-2/00851616007590/Z2460578E, Z2460647S, Z2460706A, Z2460746C, Z2460774A, Z2460865C, Z2460872B, Z2460901B, Z2460909D, Z2460909K, Z2460945H, Z2460948I, Z2460963N; NG-0730-140-2/00851616007606/Z2460578F, Z2460662A, Z2460740A , Z2460774F, Z2460850A, Z2460900B, Z2460945F, Z2460957R; NG-0740-140-2/00851616007613/Z2460578G, Z2460635C, Z2460656A, Z2460707D, Z2460733A, Z2460766A, Z2460774B, Z2460780A , Z2460865A, Z2460871A, Z2460872D, Z2460900D, Z2460909B, Z2460909I, Z2460948F, Z2460963T; NG-NV-6-30/00851616007644/Z2460776A, Z2460893A, Z2460909G, Z2460948E; NG-NV-6-40/00851616007651/Z2460776B, Z2460865D, Z2460865E, Z2460901C, Z2460909E, Z2460909J, Z2460957P, Z2460963K, Z2460963P; NG-NV-7-30/00851616007637/Z2460776C, Z2460789H, Z2460850B, Z2460900C, Z2460916A, Z2460948H, Z2460998A; NG-NV-7-40/00851616007620/Z2460776D, Z2460878A, Z2460872A, Z2460872E, Z2460901A, Z2460909C, Z2460909H, Z2460945G, Z2460948G, Z2460966B, Z2460963J;

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Neuroguard IEP 3 in →