medical Class II Updated 2018-09-05

Tecan US, Inc. recalls Cortisol Saliva Luminescence Immunoassay The IBL Interna

Recalled Product

Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.


Hazard / Issue

Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.

Issued by

FDA

Affected States: GA, IA, LA, MA, MN, OR, PA, SC, UT
Lot/Code Info: In USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132
View official government recall