medical Class II Updated 2018-09-05

Stanbio Laboratory, LP recalls Stanbio AC power adapter for the HemoPoint H2 meter, Model G

Recalled Product

Stanbio AC power adapter for the HemoPoint H2 meter, Model G3000 series. The firm name on the label for the HemoPoint H2 shows Manufactured by EKF-diagnostic GmbH, Germany.


Hazard / Issue

The US prong (type A) adapter plate may crack, break, or detach and remain in the electrical outlet exposing live electrical contacts.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial number range of affected power adapters: 41800001-41804000
View official government recall