medical Class II Updated 2018-09-05

Siemens Healthcare Diagnostics, Inc. recalls RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 106968

Recalled Product

RAPIDPoint 500 Blood Gas Analyzer; 10492730, 1069855, 10696857, 10697306 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.


Hazard / Issue

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All units affected
View official government recall