medical Class II Updated 2018-09-05

Roche Diagnostics Corporation recalls Roche cobas p 612 pre-analytical system (63x) Model: 075631

Recalled Product

Roche cobas p 612 pre-analytical system (63x) Model: 07563116001 The cobas p 612 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged and non-centrifuged sample tubes. The cobas p 612 is intended for use with analyzers that perform tests in the area of clinical chemistry, immuno chemistry, coagulation, hematology, urinalysis, nucleic acid testing.


Hazard / Issue

Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of biological material on the nozzle may cause contamination resulting in potential false positive results, depending on the sensitivity of the analytical technology.

Issued by

FDA

Affected States: CA, NY, TX, WA, WI
Lot/Code Info: All units