medical Class II Updated 2020-09-23

Terumo Cardiovascular Systems Corporation recalls Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8

Recalled Product

Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use with Terumo and Sarns brand air detection and heart-lung systems to detect gross air bubbles in the line during extracorporeal procedures.


Hazard / Issue

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI: 00886799000038; Serial Numbers: 20288 20289 20290 20291 20292 20293 20294 20295 20296 20299 20300 20301 20302 20306 20307 20308 20309 20310 20312 20317 20318 20320 20321 20322 20323 20324 20326 20328 20329 20331 20332 20334 20338 20339 20341 20342 20343 20344 20345 20346 20347 20348 20349 20350 20351 20352 20358 20365 20368 20373 20374 20375 20381 20382 20383 20384 20385 20387 20388 20392 20393 20395 20396 20397 20398 20399 20401 20402 20403 20404 20405 20408 20410 20411 20412 20555 20556 20557 20558 20559 20560 20561 20562 20563 20564 20565 20566 20567 20568 20569 20570 20571 20572 20573 20574 20575 20576 20577 20578 20579
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →