medical Class II Updated 2018-09-12

Bayer Medical Care, Inc. recalls Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT,

Recalled Product

Overhead Counterpoise Systems (OCS), OCS CONF,STD CEIL MNT, STD HORZ ARM, Catalog # OCS 115A, Material Number 59880253


Hazard / Issue

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Issued by

FDA

Affected States: OH, TX, WV
Lot/Code Info: Serial Numbers: 11051103902, 305746202