medical Class II Updated 2018-09-12

Bayer Medical Care, Inc. recalls Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUN

Recalled Product

Overhead Counterpoise Systems (OCS), OCS, 580mm CEILING MOUNT, PORTEGRA2, Catalog # OCS CEIL 58, Material Number 84433765


Hazard / Issue

The torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the system which could lead to the OCS disengaging from the ceiling mount and this may cause injury to the patient or the operator.

Issued by

FDA

Affected States: OH, TX, WV
Lot/Code Info: Serial Numbers: 917/21956, 516/21035, 817/21916, 817/21910