ArthroCare Corporation recalls Speed Stitch Needle Cassette

Recalled Product

Speed Stitch Needle Cassette


Hazard / Issue

Due to a supplier error where a portion of the raw material was inadvertently mixed. As a result, some needles may be made of commercially pure titanium as opposed to 304 stainless steel which does not meet the specification and could potentially break during use.

Issued by

FDA

Affected States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, DE
Lot/Code Info: Model: OM-8850 Lot Numbers: 2018434, 2018934, 2019558, 2019559, 2019929, 2019930, 2020931, 2022111, 2022641, 2030008, 2032212, 2033815, 2037604, 2036941, 2038701, 2039699, 2040500, 2042295, 2044957, 2047483, 2048575, 2051239 & 2032526