Siemens Medical Solutions USA, Inc recalls Software: Syngo.via RT Image Suite software version VA30_SP0
Recalled Product
Software: Syngo.via RT Image Suite software version VA30_SP02 with license T_RT_IMAGE_SUITE_PAT_MARK Intended Use: used as an advanced application on a SOM10 scanner SMN: 10496180
Hazard / Issue
Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location