Zeus Scientific, Inc. recalls ZEUS ELISA Parvovirus B19 IgM Test System, UDI 0084553300175
Recalled Product
ZEUS ELISA Parvovirus B19 IgM Test System, UDI 00845533001755 - Product Usage: intended for the qualitative detection of IgM class antibodies to human Parvovirus B19 in human serum including women of childbearing age where there is a suspicion of exposure to human Parvovirus B19.
Hazard / Issue
FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.