medical Class II Updated 2018-09-19

SPS Sterilization, Inc recalls Boston Scientific Wallstent Endoprosthesis Tracheobronchial

Recalled Product

Boston Scientific Wallstent Endoprosthesis Tracheobronchial Self-Expanding Stent, 22MM X 45MM | 75CM L, resterilized.


Hazard / Issue

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

Issued by

FDA

Affected States: AZ, CA, FL, OH, PA, TX, VA, WV
Lot/Code Info: Code Number 40451-R; Lot Number 18098774; Serial Number 21408032017-12, 21408032017-13