medical Class II Updated 2020-10-07

Ameditech Inc recalls TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-

Recalled Product

TDDA Drugs of Abuse Cup(AMT Item/ AMT Description): TDDA-6MBAU-CUP /"TDDA Drugs of Abuse Cup AMP300/BZO200/COC300/MET300/OPI300/ THC50 with Cr, Ni, pH, Bl, S.G."


Hazard / Issue

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: ALL Lots Within Expiry