Wipro GE Healthcare Private Ltd. recalls GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. I
Recalled Product
GE Giraffe Warmer, Model/Catalog Number M1118179-01090979. Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
Hazard / Issue
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Issued by
FDA