Biomerieux Inc recalls VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieu
See all recalls from Biomerieux Inc →Recalled Product
VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria
Hazard / Issue
Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.