medical Class II Updated 2024-09-25

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagno

Recalled Product

ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.


Hazard / Issue

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00630414517643 Lot Numbers: 612528, 616644, 620971, 623768, 636498, 640513, 658356, 668872