medical Class II Updated 2018-09-26

bioMerieux, Inc. recalls VITEK(R) 2 AST-P640, For susceptibility testing of Staphyloc

Recalled Product

VITEK(R) 2 AST-P640, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 418579 Product Usage: VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.


Hazard / Issue

Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 7400410103, 7400697403
View official government recall