medical Class II Updated 2018-09-26

Invivo Corporation recalls Xper Flex Cardio Physiomonitoring System Model # 45356466908

Recalled Product

Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388


Hazard / Issue

Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.

Issued by

FDA

Distribution: Distribution in the Netherlands
Lot/Code Info: System Serial Numbers: US1814000001, US1814000002, US1814000003, US1814000004, US1814000006, US1814000007, US1814000008, US1814000011, US1814000012, US1814000013, US1814000014, US1814000016, US1814000017, US1814000018, US1814000019, US1814000020, US1814000022, US1814000023, US1814000024, US1814000027, US1814000028, US1814000030, US1814000031, US1814000032, US1814000033, US1814000034, US1814000037, US1814000044, US1814000046, US1814000050, US1814000053, US1814000062, US1814000066, US1814000068,and US1814000072