medical Class III Updated 2018-09-26

Ortho-Clinical Diagnostics recalls VITROS Chemistry Products CREA Slides, Product Code 680 2584

Recalled Product

VITROS Chemistry Products CREA Slides, Product Code 680 2584 Product Usage: VITROS CREA Slides (Product Code 680 2584) - For in vitro diagnostic use only. VITROS Chemistry Products CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.


Hazard / Issue

The slides did not meet the current claims for Limit of Blank (LoB), Limit of Detection (LoD) and Limit of Quantification (LoQ) for serum and urine samples and therefore affecting the low end of the measuring range for this assay. The current low end measuring range is 0.05mg/dL for serum and 1.2mg/dL for urine. Based on the CREA testing results for the LoQ, the revised low end CREA measuring range is 0.15mg/dL for serum and 3.2mg/dL for urine.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GEN 7, GEN 10, GEN 12, GEN 13, GEN 15, GEN 16, GEN 18
View official government recall