medical Class II Updated 2024-10-02

Philips North America Llc recalls Ingenia Ambition S - For use as a diagnostic device to obtai

Recalled Product

Ingenia Ambition S - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782108 2) 781359 3) 782133


Hazard / Issue

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model UDI-DI: 782108 (01)00884838098343(21); 781359 (01)00884838090057(21). 782133 N/A Serial Numbers: 27506 27508 27509 27512 27514 27515 27516 27517 27518 27519 27520 27523 27524 27525 27526 27528 27531 48146 48164 48550 48556 48588 48589 48608 48612 48627 48648 48661 48662 48663 48664 48665 48666 48667 48669 48672 48679 48686 48688 48694 48696 48697 48698 48699 48700 48707 48708 48710 48714 48715 48721 48725 48726 48728 48729 48738 48744 48749 48754 48800 48804 48810 48813 48833
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →