Howmedica Osteonics Corp. recalls ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR
See all recalls from Howmedica Osteonics Corp. →Recalled Product
ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORPIO + MOBILE BEARING PCS 78-5-0910 SCORPIO C-DOME PATELLA 73-2710 SCORPIO INSET PATELLA 73-2610 SCORPIO M-DOME PATELLA 73-0110 73-0710 73-0910 SCORPIO RECESSED PATELLA 26MM 3044-0026 SCORPIO TS TIB INS 72-4-9114 SCORPIO TS TIBIAL INSERT 72-4-0324 SCORPIOFLEX TOTAL KNEE CR 72-16-0512 72-16-0518 72-16-0912 72-16-1112 SERIES II TIBIAL BEAR INSERT NO 11 15MM 3052-1115 UHR BIPOLAR 26X52mm UH1-52-26 UHR BIPOLAR 26X58mm UH1-58-26 UHR BIPOLAR 28X48mm UH1-48-28 UHR BIPOLAR 28X52mm UH1-52-28 UHR BIPOLAR 28X55mm UH1-55-28 UHR BIPOLAR 32X49MM UH1-49-32 UHR BIPOLAR 32X51MM UH1-51-32 UHR BIPOLAR 32X52MM UH1-52-32 UHR BIPOLAR 32X54MM UH1-54-32 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Hazard / Issue
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Issued by
FDA
✅ What Should You Do?
- 1Contact your healthcare provider immediately to discuss this recall.
- 2Stop using the device if it is safe to do so, and contact your provider.
- 3Follow the manufacturer's remedy instructions provided in the recall notice.
- 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
Looking for a safe alternative?
Search Amazon for ALL POLY CONSTRAINED INSERT →If you were harmed by this recalled product, Top Class Actions lists active class action lawsuits and settlements that may apply to you.
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.