medical MODERATE Updated 2026-09-23

Howmedica Osteonics Corp. recalls TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIA

See all recalls from Howmedica Osteonics Corp. →

Recalled Product

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for TRIATHLON HINGE INSERT SIZE →
View official government recall