medical HIGH RISK Updated 2026-10-07

Medline Industries, LP recalls Medline Basic Homecare Beds and Medlite Homecare Beds and Ac

See all recalls from Medline Industries, LP →

Recalled Product

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: MDR107000P UDI-DI 00080196288284 All lots

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Medline Basic Homecare Beds →
View official government recall