medical Class II Updated 2024-10-02

American Contract Systems, Inc. recalls Open Shoulder, LWOS39L; Medical convenience kit

Recalled Product

Open Shoulder, LWOS39L; Medical convenience kit


Hazard / Issue

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00191072177032; Lot: 2405141; Exp: 05/14/2025