medical Class II Updated 2018-10-03

Stryker GmbH recalls Hoffmann LRF Hexapod Software Catalog #: 983031 Softwa

Recalled Product

Hoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002


Hazard / Issue

When creating a correction plan with the Correct Axial First button selected and the No. of corrections per day set to more than 1 (e.g., 2, 3, or 4), the Hexapod Software computes an erroneously accelerated correction plan for the axial portion of the plan.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
View official government recall