medical MODERATE Updated 2026-10-07

Alere San Diego, Inc. recalls McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5

See all recalls from Alere San Diego, Inc. →

Recalled Product

McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/Urine¿hCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 5001/ 20612479202744/1155958203, 12/15; 92210/10815845020130/1156015803, 07/16; 23900530/00613647000571/1155810105, 2018-01-15; 23900531/00613647000533/1155810105, 2018-01-15; 5002/20612479202768/1155958305, 12/15; 92215/10815845020123/1156015903, 07/16; B-1077-23/20885380020268/1155904805, 9/2014; MPH22025HCG/10884389191880/1156093402, 04-01-2015

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for McKesson CONSULT hCG Urine →
View official government recall