Home
medical
Z-3310-2026
medical
MODERATE
Updated 2026-10-07
Philips North America recalls Ingenia Ambition S and Ingenia Ambition X.; Model number (R
See all recalls from Philips North America →
Recalled Product
Ingenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 782159 (S), 5. 782194 (S), 6. 781356 (X), 7. 782109 (X), 8. 782138 (X), 9. 782160 (X), 10. 782195 (X); Software Versions: R11, R12 and R13;
Hazard / Issue
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Distribution: Nationwide (United States)
Lot/Code Info: 1. Model number (REF): 781359 (S); UDI-DI: 00884838090057; Serial Numbers: 48010, 48023, 48028, 48047, 48122, 48160, 48173, 48180, 48196, 48286; 2. Model Number (REF): 782108 (S); UDI-DI: 00884838098343; Serial Numbers: 48507, 48510, 48530, 48548, 48554, 48576, 48577, 48580, 48586, 48607, 48616, 48620, 48653, 48686, 48688, 48726, 48746, 48778, 48818, 48831, 48847, 48872, 48878, 48888, 48893, 48896, 48897, 48917, 48922, 48931; 3. Model Number (REF): 782139 (S); UDI-DI: 00884838108639; Serial Numbers: 29016, 29017, 29018, 29020, 29026, 29029, 29031, 29041, 29054, 29063, 29069, 29071, 29076, 29078, 29079, 29086, 29104, 29115, 29118, 29144, 29145, 29150, 29160, 29164, 29169; 4. Model Number (REF): 782159 (S); UDI-DI: 00884838115705; Serial Numbers: 91007, 91027, 91031, 91038, 91043, 91051; 5. Model Number (REF): 782194 (S); UDI-DI: 00884838128477; Serial Numbers: 22401, 22410; 6. Model Number (REF): 781356 (X); UDI-DI: 00884838090040; Serial Numbers: 47004, 47006, 47011, 47012, 47013, 47014, 47018, 47020, 47026, 47027, 47028, 47029, 47037, 47038, 47039, 47043, 47048, 47051, 47057, 47059, 47063, 47067, 47068, 47070, 47092, 47094, 47098, 47099, 47177, 47187, 47188, 47189, 47195, 47197, 47199, 47203, 47206, 47211, 47215, 47221, 47222, 47233, 47237, 47242, 47248, 47253, 47260, 47268, 47270, 47271, 47274, 47276, 47278, 47279, 47281, 47282, 47289, 47290, 47292, 47294, 47297, 47303, 47305, 47313, 47314, 47322, 47344, 47347, 47348, 47351, 47353, 47356, 47371; 7. Model Number (REF): 782109 (X); UDI-DI: 00884838098350; Serial Numbers: 47502, 47503, 47505, 47507, 47508, 47512, 47517, 47518, 47519, 47520, 47534, 47535, 47541, 47542, 47543, 47549, 47552, 47555, 47558, 47562, 47567, 47568, 47569, 47570, 47572, 47580, 47588, 47589, 47590, 47594, 47595, 47596, 47598, 47599, 47608, 47610, 47612, 47614, 47615, 47622, 47623, 47634, 47642, 47646, 47648, 47651, 47653, 47654, 47661, 47662, 47663, 47668, 47669, 47671, 47674, 47681, 47682, 47687, 47692, 47693, 47694, 47699, 47700, 47702, 47703, 47706, 47709, 47710, 47711, 47714, 47717, 47718, 47719, 47721, 47722, 47723, 47724, 47726, 47727, 47728, 47729, 47730, 47731, 47732, 47733, 47752, 47760, 47761, 47762, 47766, 47767, 47771, 47778, 47779, 47780, 47781, 47788, 47789, 47795, 47796, 47797, 47800, 47807, 47808, 47809, 47812, 47815, 47816, 47817, 47818, 47820, 47823, 47824, 47826, 47834, 47839, 47843, 47848, 47849, 47851, 47862, 47863, 47864, 47874, 47876, 47881, 47882, 47884, 47885, 47886, 47887, 47891, 47892, 47893, 47894, 47895, 47899, 47932, 47944, 47962; 8. Model Number (REF): 782138 (X); UDI-DI: 00884838108622; Serial Numbers: 49000, 49001, 49004, 49014, 49015, 49019, 49020, 49022, 49025, 49027, 49028, 49033, 49037, 49038, 49041, 49043, 49045, 49047, 49048, 49049, 49053, 49056, 49057, 49058, 49062, 49063, 49071, 49072, 49075, 49097, 49099, 49100, 49102, 49106, 49109, 49111, 49113, 49124, 49128, 49129, 49136, 49145, 49153, 49154, 49155, 49156, 49158, 49161, 49162, 49163, 49164, 49165, 49168, 49170, 49171, 49172, 49174, 49177, 49178, 49180, 49183, 49189, 49190, 49194, 49196, 49197, 49203, 49206, 49207, 49210, 49212, 49215, 49216, 49217, 49219, 49220, 49221, 49222, 49223, 49228, 49229, 49230, 49231, 49233, 49237, 49238, 49243, 49246, 49247, 49250, 49251, 49253, 49257, 49261, 49265, 49268, 49269, 49270, 49272, 49282, 49287, 49288, 49290, 49291; 9. Model Number (REF): 782160 (X); UDI-DI: 00884838115712; Serial Numbers: 92000, 92003, 92006, 92016, 92022, 92024, 92028, 92029, 92031, 92035, 92036, 92039, 92042, 92046, 92053, 92056, 92067, 92068, 92069, 92070, 92072, 92084, 92097; 10. Model Number (REF): 782195 (X); UDI-DI: 00884838128484; Serial Numbers: 55004, 55005, 55011, 55012, 55018, 55019, 55021, 55024, 55032;
✅ What Should You Do?
1 Contact your healthcare provider immediately to discuss this recall. 2 Stop using the device if it is safe to do so, and contact your provider. 3 Follow the manufacturer's remedy instructions provided in the recall notice. 4 Report device problems to FDA MedWatch or call 1-800-FDA-1088.
If you were harmed by this recalled product, Top Class Actions lists active class action lawsuits and settlements that may apply to you.
View official government recall
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime