Brahms GmbH Recalls

1 recall on record in the United States

medical: 1
medical Class II
2026-04-15

Brahms GmbH recalls B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in

Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.