Brahms GmbH recalls B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in
See all recalls from Brahms GmbH →Recalled Product
B.R.A.H.M.S PlGF Plus KRYPTOR, Catalog Number: 859075N. in vitro diagnostic test for Placental Growth Factor
Hazard / Issue
Customers have observed quality control values out of range at the lowest quality control level (QC Level 1) for the PIGF assay which require troubleshooting steps to resolve. If Quality Controls are not performed according to the Instructions for Use, the underestimation of QC1 may go undetected and could contribute to falsely elevated sFlt-1/PlGF ratio results. A falsely elevated sFlt-1/PlGF ratio may result in a higher-risk classification for progression to preeclampsia with severe features potentially leading to intensified clinical monitoring.
Issued by
FDA
✅ What Should You Do?
- 1Contact your healthcare provider immediately to discuss this recall.
- 2Stop using the device if it is safe to do so, and contact your provider.
- 3Follow the manufacturer's remedy instructions provided in the recall notice.
- 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
Looking for a safe alternative?
Search Amazon for B.R.A.H.M.S PlGF Plus KRYPTOR, →If you were harmed by this recalled product, Top Class Actions lists active class action lawsuits and settlements that may apply to you.
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime