1 recall on record in the United States
Emano Metrics, Inc. recalls Emano Flow with software version v2.0.0 GS1 GTIN code: 1998
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.