medical MODERATE Updated 2026-07-29

Emano Metrics, Inc. recalls Emano Flow with software version v2.0.0 GS1 GTIN code: 1998

See all recalls from Emano Metrics, Inc. →

Recalled Product

Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Emano Flow with software →
View official government recall