medical MODERATE Updated 2026-04-08

Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Pr

See all recalls from Olympus Corporation of the Americas →

Recalled Product

Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Laser System: TFL-SLS; UDI: 00821925044135; Component Part: TFL-CSLU; UDI: 00821925044593; Serial #: MDUF190114, MDUF220073, MDUF220331, MDUF220353, MDUF220354, MDUF220356, MDUF220411, MDUF220412, MDUF220413, MDUF220414, MDUF220427, MDUF220451, MDUF220453, MDUF220454, MDUF220455, MDUF220456, MDUF230004, MDUF230005, MDUF230006, MDUF230007, MDUF230076, MDUF230140, MDUF230141, MDUF230163, MDUF230164, MDUF230166, MDUF230167, MDUF230234, MDUF230283, MDUF230284, MDUF230331, MDUF230332, MDUF230361, MDUF230362, MDUF230363, MDUF230364, MDUF230365, MDUF230366, MDUF230411, MDUF230412, MDUF230413;

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Brand Name: Olympus SOLTIVE →
View official government recall