medical MODERATE Updated 2026-08-26

Tobii Dynavox Llc recalls TD Navio Maxi Speech Generating Device Model/Catalog Number

See all recalls from Tobii Dynavox Llc →

Recalled Product

TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for TD Navio Maxi Speech →