medical MODERATE Updated 2026-08-26

Beckman Coulter Inc. recalls Beckman Coulter DxC 500i Clinical Analyzer Modules using sof

See all recalls from Beckman Coulter Inc. →

Recalled Product

Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 1) Part Number C63522, UDI/DI 14987666545089; 2) Part Number C63521, UDI/DI 14987666545072; 3) Part Number C13252, UDI/DI 15099590742331.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Beckman Coulter DxC 500i →
View official government recall