medical MODERATE Updated 2026-08-26

Medtronic MiniMed, Inc. recalls MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW

See all recalls from Medtronic MiniMed, Inc. →

Recalled Product

MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

Issued by

FDA

Distribution: International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Lot/Code Info: Model/UDI-DI Serial Numbers: MMT-1885CLW/00763000442187, 00763000974374/NG3150737H, NG3155194H, NG3151561H, NG3154693H, NG3151716H, NG3153547H, NG3153637H, NG3155407H, NG3152666H, NG3155788H, NG3081975H, NG2901117H, NG3080842H, NG3081463H, NG3153416H, NG3081894H, NG3153781H, NG3155944H, NG3081982H, NG3081525H, NG3155909H, NG3090722H, NG3154897H, NG3088436H, NG3088708H, NG3090073H, NG3155908H, NG3150720H, NG3092957H, NG3128715H, NG3128784H, NG3152774H, NG3153929H, NG3514566H, NG3150780H, NG3151290H, NG3151508H, NG3507161H, NG3508693H, NG3081976H, NG3151460H, NG3152051H, NG3152085H, NG3152344H, NG3153213H, NG3152772H, NG3155586H, NG3150786H, NG3153214H, NG3153385H, NG3151840H, NG3082019H, NG3153405H, NG3153788H, NG3153908H, NG3154527H, NG3151598H, NG2854971H, NG3155196H, NG3155202H, NG3155341H, NG3091491H, NG3155557H, NG3081498H NG3155688H, NG3155020H, NG3508565H, NG3155419H, NG3507536H, NG3091563H, NG3082036H, NG3154689H, NG3154995H, NG3508504H, NG3514345H, NG3509850H, NG3527505H, NG3514547H, NG3527469H, NG3504333H, NG3504598H, NG3505071H, NG3505220H, NG3505273H, NG3514617H, NG3506727H, NG3085044H, NG3507542H, NG3507547H, NG3514448H, NG3508303H, NG3508350H, NG3508371H, NG3514612H, NG3508511H, NG3508617H, NG3508698H, NG3508707H, NG3508757H, NG3508799H, NG3508903H, NG3509138H, NG3509295H, NG3509299H, NG3509452H, NG3509685H, NG3509689H, NG3509836H, NG3509914H, NG3509925H, NG3514375H, NG3514441H, NG3505246H, NG3505166H, NG3505283H, NG3151570H, NG3154948H, NG3514667H, NG3527165H, NG3527243H, NG3527298H, NG3527468H, NG3527480H, NG3527491H, NG3527580H, NG3527529H, NG3527543H, NG3527571H, NG3527574H, NG3092722H, NG3527581H, NG3527611H, NG3527637H, NG3527642H, NG3527672H, NG3527710H, NG3527737H, NG3527760H, NG3527865H, NG3527918H, NG3527961H, NG3527989H, NG3528017H, NG3528023H, NG3528036H, NG3528061H, NG3528120H, NG3528147H, NG3528156H, NG3528158H, NG3528180H, NG3528218H, NG3528226H, NG3528378H, NG3539964H, NG3539969H, NG3539970H, NG3539999H, NG3540001H, NG3540013H, NG3540029H, NG3540046H, NG3540053H, NG3540084H, NG3540191H, NG3540270H, NG3540310H, NG3540320H, NG3540581H, NG3540604H, NG3540628H, NG3540629H MMT-1886CLW/00763000442194, 00763000974381/NG2966956H, NG2970938H, NG2986362H, NG3013390H, NG3057910H, NG3167493H, NG3466128H, NG3098147H, NG3136398H, NG3137113H, NG3139432H, NG3143920H, NG3144721H, NG3145486H, NG3145488H, NG3471528H, NG3167609H, NG3147120H, NG3147035H, NG3430059H, NG3438451H, NG3464758H, NG3163796H, NG3166778H, NG3409587H, NG3167450H, NG3167623H, NG3167967H, NG3169694H, NG3170253H, NG3321081H, NG3409583H, NG3409611H, NG3409656H, NG3409680H, NG3409697H, NG3409716H, NG3409807H, NG3409825H, NG3409831H, NG3409854H, NG3409925H, NG3409959H, NG3410005H, NG3410053H, NG3410092H, NG3410111H, NG3410166H, NG3410189H, NG3410227H, NG3410233H, NG3410243H, NG3410279H, NG3410304H, NG3410389H, NG3410398H, NG3429103H, NG3429377H, NG3429485H, NG3429881H, NG3430107H, NG3430448H, NG3437945H, NG3437957H, NG3438394H, NG3470788H, NG3466185H, NG3466619H, NG3466620H, NG3467368H, NG3467586H, NG3468759H, NG3469430H, NG3469907H, NG3470728H, NG3168104H, NG3472157H, NG3169669H, NG3170079H, NG3430111H, NG3429099H, NG3437621H, NG3167969H, NG3465813H, NG3167397H

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for MiniMed 780G Insulin Pump, →
View official government recall